• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number CNWTT0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Type of Reportable Event Serious Injury
Event or Problem Description
A nurse reported following an intraocular lens implant procedure, patient didn't feel the good vision at far, intermediate, and near distances.The cause of the non-improved visual acuity could not be found, it was judged that the patient could not adapt to the multifocal iol, and exchanged with the monofocal iol.Additional information has been requested.There are two medical device reports associated with this event.This report is associated with the right eye.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint could not be confirmed.Iol received adhered to gauze in a zip lock bag.Solution is dried on both surfaces of the optic and both haptics.Both haptics are rejectable with fibers.The optic is cut/split in two and scratched/marked-rejectable with fibers.The cause of the non-improved visual acuity could not be found, it was judged that the patient could not adapt to the multifocal iol, and exchanged with the monofocal iol.¿ the root cause for the reported complaint could not be determined.As per the reporter, it was suggested that the patient's ability to adapt to multifocal iol could be a contributing factor.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14772767
Report Number9612169-2022-00289
Device Sequence Number12300431
Product Code MFK
UDI-Device Identifier00380652408059
UDI-Public00380652408059
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/18/2024
Device Model NumberCNWTT0
Device Catalogue NumberCNWTT0.165
Device Lot Number25242367
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-