• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND GLENOSPHERE ECC D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND GLENOSPHERE ECC D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE Back to Search Results
Model Number 130760042
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Unspecified Tissue Injury (4559)
Date of Event 06/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Left shoulder replacement revision for dislocation - surgeon noted that in primary shoulder (completed by him (b)(6) 2022) the metaglene has been positioned quite high on the face of the glenoid, causing the humeral cup to notch on the inferior neck of glenoid.Surgeon removed glenosphere and humeral cup, removed extra bone from neck of glenoid and replaced the glenosphere as well as adding a retentive cup doi: (b)(6) 2022.Dor: (b)(6) 2022.Left shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: device associated with this report was not received for examination.All available x-rays and photographs were reviewed.A complaint is confirmed, a dislocation event can be observed.However, no product defects that could have contributed to the reported event were identified.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the metaglene was mispositioned.No delay reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE ECC D42MM
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14773036
Report Number1818910-2022-11406
Device Sequence Number1294215
Product Code HSD
UDI-Device Identifier10603295027751
UDI-Public10603295027751
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K120174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number130760042
Device Catalogue Number130760042
Device Lot Number5389754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2022
Supplement Date Received by Manufacturer08/26/2022
08/30/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date08/29/2022
08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND GLENOSPHERE ECC D42MM; DXTEND STAND PE CUP D42 +6MM; DXTEND STAND PE CUP D42 +6MM; METAGLENE; UNKNOWN SHOULDER METAGLENE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-