| Model Number |
1217-31-056 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Pinnacle first implantation (b)(6) 2022 too medial.Was surgery delayed due to the reported event? --> unknown, was procedure successfully completed? --> unknown, were fragments generated? --> unknown, if yes, were they removed easily without additional intervention? --> unknown, patient status/ outcome / consequences --> , was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, ip-(b)(6).Device property of -->none, device in possession of -->none, ip-(b)(6).Device property of -->none, device in possession of -->none, ip-(b)(6).Device property of -->none, device in possession of -->none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that there was no surgical delay reported and the reason for the removal surgery was the malposition of the implants anteversion was bad.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Visual analysis of the x-ray photos were reviewed.Based on the provided evidence, it cannot be observed a misposition of the device.Neither a defect or malfunction can be established that could contribute to the reported event, therefore, the allegation is not confirmed.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
|
| |
|
Search Alerts/Recalls
|