The complaint device was not to be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.The production site checked the batch record of impacted monoject hypodermic safety needle 25g x 5/8", ref 1182558, lot: 02102036 with no anomalies found on batch documentation.No samples were received by the customer for evaluation.Archive sample evaluation: 50pcs of archive samples have been visually checked with no anomalies found on the needle hub.5pcs of archive samples have been checked in accordance with iso 80369-20:2015, annex e.All samples tested passed the test.Risk statement: the overall risk assessment, given the severity assessment of s3-moderate injury and a probability assessment of p2-low, the final risk priority number is 6.Risk level is in green area/low risk according to our risk management procedure.
|