| Model Number |
15 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that non of the buttons on their device are working aside from the on/off button.As a result, defibrillation therapy may be unavailable, if needed.The buttons started working again after the device was power cycled multiple times.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker replaced the device's main keypad assembly and system pcb flex cable as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|