Manufacturing and quality control data: the progav® shunt system was manufactured by a qualified employee in march 2013.Deviations during assembly did not occur.The system was sterilized by miethke and released for shipment after final inspection.For the purposes of this investigation, the progav® shunt system is comprised of a progav® adjustable pressure valve with a normal pressure range of 0 to 20 cmh2o and a shuntassistant® fixed pressure valve with a fixed pressure range of 25 cmh2o.The shunt system was inspected as article fv421t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Visual inspection: a deformation of the outer housing of the progav valve and depositis was observed through the visual inspection.The measurement of the plane parallelism could confirm that with a value of -0,101 mm the housing membrane is clearly outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the shuntassistant® is operating within the specified tolerances in the vertical position.Adjustment test: the progav® was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function did not operate as expected.The results indicate that the rotor no longer performs as required.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: it should be noted that the product was received dry (i.E.Not submersed in liquid as recommended).The investigation of dry items is not significant due to the affect dry deposits of liquor and blood can have on product performance.In spite of this, we have investigated the valve to the best of our abilities.Based on our investigation, we confirm that the valve was non-adjustable at the time of our investigation.The cause of the deformation of the progav® valve and the resultant defect of the rotor could not be determined through our investigation.Significant outside pressure, for example by too much force from the progav® adjustment tool or by a fall or impact to the head of the patient, can compromise the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
|
It was reported that a progav shunt system(#fv421t) was implanted during a procedure performed on (b)(6) 2013.According to the complainant, the shunt system was believed to be operated in underdrainage and to be not adjustable.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 9 years, height: 130 centimeters (cm), weight: 41 kilograms (kg), gender: male.
|