• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PRECHAMBER PED.; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PRECHAMBER PED.; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV035T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a pediatric prechamber (#fv035t) was implanted during a procedure performed on (b)(6) 2020.According to the complainant, the shunt system was believed to be blocked.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 6 years, height: 89 centimeters (cm), weight: 10.5 kilograms (kg), gender: female.Same event as medwatch-no.3004721439-2022-00173.
 
Additional Manufacturer Narrative
Investigation: visual inspection: deposits and bloody residues in the pediatric prechamber, damaged third-party catheter.Permeability test: the test showed that the pediatric prechamber with the progav 2.0 shunt system is permeable.During the test, bloody liquid leaked from the shunt system.Results: for the sake of thoroughness and transparency, we would like to point out that a permeability test was already carried out after the revision at the hospital.Based on our investigation results, we can detect deposits.At the time of the investigation, the deposits do not affect the technical properties of the pediatric prechamber.We are unable to determine how the aforementioned functional impairment occurred at the time of the investigation.We can exclude a defect at the time of release.The pediatric prechamber met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECHAMBER PED.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14775442
Report Number3004721439-2022-00174
Device Sequence Number1349265
Product Code JXG
UDI-Device Identifier04041906117603
UDI-Public4041906117603
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K011030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2025
Device Model NumberFV035T
Device Catalogue NumberFV035T
Device Lot Number20048632
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Initial Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexFemale
Patient Weight10 KG
-
-