• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*CUTTER VASCULAR 55MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*CUTTER VASCULAR 55MM; STAPLE, IMPLANTABLE Back to Search Results
Model Number TVC55
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
See attached user facility medwatch.
 
Additional Manufacturer Narrative
(b)(4).Batch # unk.User facility medwatch # (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Additional Manufacturer Narrative
(b)(4) date sent: 07/18/2022 d4: batch # 430a83 additional information was requested, and the following was received: could you please clarify if there were any patient consequences? staple line had to be oversewn with suture, no other patient consequences could you please clarify if was any change in the post-operative care of the patient as a result of the event? unknown investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tvc55 device was received with no apparent damage and with a reload loaded in the device.The reload was returned fully fired.The device was tested for functionality with a test reload and it fired, cut, and formed all the staples as intended.The staple line and cut line were complete, and the staples meet the staple form release criteria.In addition, a tyvek was received along with the device.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: ensure that the tissue lies flat between the forks.Any ¿bunching¿ of tissue along the reload or scale may result in an incomplete staple line.Tissue to be transected must be located between the arrows marked on the instrument jaw.Any tissue located outside of the arrows is out of the stapling range.The event described could not be confirmed as the device performed without any difficulties.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
4188 lexington avenue north
shoreview 55126
MDR Report Key14775668
Report Number3005075853-2022-04039
Device Sequence Number11100191
Product Code GDW
UDI-Device Identifier10705036013389
UDI-Public10705036013389
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/16/2026
Device Model NumberTVC55
Device Catalogue NumberTVC55
Device Lot Number470A89
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-