| Model Number |
110-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Granuloma (1876)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient's guardian reported "rupture" against left side device.The event of "injection granulomas" was reported via mri report.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of granuloma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and "injection granulomas.".
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data.Device evaluation: the device related to the reported event of rupture and granuloma was received on august 01, 2022, with lot number: 2747318.Analysis of the returned device identified: opening on posterior, blue particles in the shell and underweight.A visual and microscopic analysis was performed which identified: sharp opening on posterior, crease fold, wear abrasion and stress marks.Base on the device analysis the final assessment is: a sharp pattern on posterior of opening device assessed as a surgical impact opening, for the stress marks in the shell.
|
| |
|
Event or Problem Description
|
|
Patient's guardian reported "rupture" against left side device.The event of "injection granulomas" was reported via mri report.Device has been explanted.
|
| |
|