|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The reported patient effect of dissection is listed in the xience skypoint everolimus eluting coronary stent systems instructions for use (ifu) as a known patient effect of coronary stenting procedures.The investigation determined the reported failure to advance appears to be related to the operational context of the procedure.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a tortuous right coronary artery.Three xience skypoint stent delivery systems (2.5x15mm, 2.5x12mm, and 2.5x08mm) failed to cross the lesion; however, a dissection was noted.It is unknown if there was any treatment provided for the dissection.An unspecified stent was used to successfully complete the procedure.There was no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
|