| Model Number |
A2020 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
05/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported that the tip on the mayfield radiolucent skull pins (a2020) splintered and broke off in the patient's scalp, and the surgeon needed to explore the scalp to find the lost piece.The incident occurred when the pins were being removed at the end of craniotomy procedure; this caused quite a bit of bleeding which ended up making the case go longer for 20 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
The mayfield radiolucent skull pins (a2020) was returned for evaluation: failure analysis - complaint confirmed via inspection of the returned skull pin.The tip of the pin was broken and part of the pin was missing.Root cause - probable root cause includes damage to the pin from mishandling (such as dropping the skull pin) or from improper seating of the skull pin in the mayfield skull clamp.
|
| |
|