• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that the 3 lead-electrodes of their lifepak 20 device had burned the skin of the patient.It was reported that the patient did sustain skin burns.The patient associated with the reported event did not survive, however the health care provider involved stated that they do not believe the device use contributed to the patient outcome.
 
Additional Manufacturer Narrative
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.A clinical review was performed and determined that there is insufficient data available to determine the impact of the device use, and therefore it is concluded the device contribution to the reported event is unknown.Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker will not request any patient identifying information to be in accordance with regulation (eu) (b)(4) of the (b)(6) parliament and of the council.
 
Additional Manufacturer Narrative
Stryker has made multiple attempts to contact the customer regarding the status of their device, but no response has been received from the customer.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that the 3 lead-electrodes of their lifepak 20 device had burned the skin of the patient.It was reported that the patient did sustain skin burns.The patient associated with the reported event did not survive, however the health care provider involved stated that they do not believe the device use contributed to the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key14777449
Report Number0003015876-2022-01259
Device Sequence Number5936328
Product Code MKJ
UDI-Device Identifier00883873832622
UDI-Public00883873832622
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20
Device Catalogue Number99402-000048
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age23 WK
Patient SexUnknown
-
-