| Model Number |
2112-90006 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A company representative reported that a capri inserter, capri inserter inner shaft, and capri non-tapered driver were damaged during surgery.The surgeon malleted the inserter into the patient which caused deformation on the proximal end of the capri inserter (mallet marks) and deformation on the distal end of the capri inserter inner shaft.Images provided indicate the distal end of the capri non-tapered driver is also deformed.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.This record captures the capri inserter.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the device and the reported issue of inserter damage was confirmed.The proximal end of the inserter exhibits severe mallet marks indicating that excessive force was applied on the inserter during implant insertion.A review of complaint history associated with the subject catalog and lot number was performed, and no adverse trend was observed.The device was inspected prior to the procedure and no damage was noticed.Hence, the cause of the reported event is application of excessive force by the user causing device deformation.
|
| |
|
Event or Problem Description
|
|
A company representative reported that a capri inserter, capri inserter inner shaft, and capri non-tapered driver were damaged during surgery.The surgeon malleted the inserter into the patient which caused deformation on the proximal end of the capri inserter (mallet marks) and deformation on the distal end of the capri inserter inner shaft.Images provided indicate the distal end of the capri non-tapered driver is also deformed.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.This record captures the capri inserter.
|
| |
|
Search Alerts/Recalls
|