| Model Number |
12-30-0348 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Returned device was visually inspected, functionally tested and the complaint was confirmed.The device left locking mechanism/cam had disassembled from the plate.The device meets dimensional specifications.The cam is not marred/bent/damaged from over-rotation or being in contact with a proud screw.Unknown factors include: if the locking mechanisms were properly locked at time of implantation, if the cam was damaged/loosened as a result of handling in the field, patient activity at the time or prior to the event, patient bone quality, the degree of spinal instability or patient compliance with post-operative care instructions or if the patient sustained a fall/impact of any sort (no trauma was reported).Root cause is unknown.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, patient underwent anterior cervical fusion (acdf) surgery with a three level anterior cervical plate (acp) c3-c6.During routine follow-up, radiographs depict the left c6 locking mechanism on the acp had disassociated.The bone screw has backed out 2-3 mm.Revision was performed on june 7, 2022, to replace the plate.
|
| |
|