The reported event was inconclusive as no sample was returned for evaluation.It was unknown whether the device had met relevant specifications.A potential root cause for this failure mode could be ¿short latex dwell time" or "latex dip speed out too slow" and might be contributed by user related (example: contact with sharp object)/ exposure to petrolatum based products mechanical failure/operator error).The device was not returned for evaluation.The lot number is unknown.Therefore, the device history record could not be reviewed.The instructions for use were found adequate and states the following: "a review of the device labeling have been conducted for investigation purposed.Warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient".
|