• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number RA002-4545SL70
Medical Device Problem Codes Corroded (1131); Structural Problem (2506); Mechanical Jam (2983); Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to loss of extension.During examination of the rod, the retrieval center found there was wear, corrosion and other surface damage.Additionally, debris was found inside the rod and it was jammed.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.
 
Additional Manufacturer Narrative
Complaint 1 of 2.No product has been returned for evaluation.Evaluation was performed by london implant retrieval centre (lirc).The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.A review of the dhr documents indicated that the rod was manufactured by the specified requirements and met all of the required quality inspections prior to shipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
8583448112
MDR Report Key14779429
Report Number3006179046-2022-00209
Device Sequence Number17587414
Product Code PGN
UDI-Device Identifier00856719002244
UDI-Public856719002244
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K161751
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRA002-4545SL70
Device Lot NumberA160210-16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2022
Initial Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/10/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age8 YR
Patient SexMale
Patient Weight22 KG
-
-