| Model Number |
MS1-4570R |
| Medical Device Problem Codes |
Corroded (1131); Structural Problem (2506); Mechanical Jam (2983); Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint 2 of 2.No product has been returned for evaluation.Evaluation was performed by (b)(4).The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.A review of the attached dhr documents indicated that the rod was manufactured by the specified requirements and met all of the required quality inspections prior to shipment.
|
| |
|
Event or Problem Description
|
|
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to loss of extension.During examination of the rod, the retrieval center found wear, galling, corrosion, scratching, discoloration and other surface damage.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.
|
| |
|