| Model Number |
A34-34/C100V |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Stenosis (2263)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft, an afx vela infrarenal, an afx limb stent graft, and an ovation ix extender.This procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately (4) months post-initial procedure, a non-device related reintervention was performed with the implant of three (3) ovation ix iliac limbs.Approximately seven (7) months later, a non-device related reintervention was performed with the implant of an ovation ix iliac limb and an afx limb stent graft.Approximately three (3) months later, a non-device related reintervention was performed with the implant of an afx limb stent graft and two (2) ovation ix iliac limbs.Now, approximately one (1) year later, a possible type ia endoleak was identified at routine follow-up.Reportedly, the physician planned to treat the patient with a non-endologix device on 06/13/2022.It is unknown if the reintervention took place.The patient status is not known.
|
| |
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
|
| |
|
Event or Problem Description
|
|
The patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft, an afx vela infrarenal, an afx limb stent graft, and an ovation ix extender.This procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately (4) months post-initial procedure, a non-device related reintervention was performed with the implant of three (3) ovation ix iliac limbs.Approximately seven (7) months later, a non-device related reintervention was performed with the implant of an ovation ix iliac limb and an afx limb stent graft.Approximately three (3) months later, a non-device related reintervention was performed with the implant of an afx limb stent graft and two (2) ovation ix iliac limbs.Now, approximately one (1) year later, a possible type ia endoleak was identified at routine follow-up.Reportedly, the physician planned to treat the patient with a non-endologix device on (b)(6) 2022.It is unknown if the reintervention took place.The patient status is not known.Clarification: the patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft, an afx vela infrarenal, an afx limb stent graft, and an ovation ix extender.This procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately (4) months post-initial procedure, a non-device related reintervention was performed with the implant of one (1) ovation ix iliac limb and two (2) afx limb stent grafts.Approximately seven (7) months later, a non-device related reintervention was performed with the implant of an ovation ix iliac limb and an afx limb stent graft.Approximately two(2) months later, a non-device related reintervention was performed with the implant of an afx limb stent graft and three (3) ovation ix iliac limbs.Now, approximately one (1) year later, a possible type ia endoleak was identified at routine follow-up.Reportedly, the physician planned to treat the patient with a non-endologix device on (b)(6) 2022.It is unknown if the reintervention took place.The patient status is not known.Additional information: clinical assessment confirmed that the patient was treated (as planned) on (b)(6) 2022.Reportedly, the patient was discharged home two days postoperative.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type ia endoleak is unconfirmed.The planned additional endovascular procedure dated (b)(6) 2022 is confirmed as completed.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported as being discharged on the second post operative day home.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.Corrections: b2 outcomes attributed to ae.B5 describe event or problem.E1 initial reporter.G3 awareness date.H6 health effect - impact code; remove 4614.H6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type ia endoleak and additional endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest the left iliac artery was poorly perfused (not occluded, but stenosed), which was not included in the event as reported.This finding was discovered during an examination of the angiogram dated 13 june 2022.The nine stents implanted in the left iliac from index to 13 june 2020 (cautionary product use) could have contributed to the poor left iliac perfusion; in addition, these nine stents were a combination of ovation and afx limbs (off-label condition).The infrarenal vela extension was also noted to be 23mm below the left renal artery per ct (computed tomography) scan dated 19 december 2020 and angiogram dated 13 june 2020; it is unclear if this was a migration of the stent graft or suboptimal placement at the index procedure.The position of the infrarenal vela extension could have contributed to the reported type ia endoleak.Device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.No procedure-related harms were identified.The final patient status was reported as being discharged on the second post operative day.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.
|
| |
|
Event or Problem Description
|
|
The patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft, an afx vela infrarenal, an afx limb stent graft, and an ovation ix extender.This procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately (4) months post-initial procedure, a non-device related reintervention was performed with the implant of three (3) ovation ix iliac limbs.Approximately seven (7) months later, a non-device related reintervention was performed with the implant of an ovation ix iliac limb and an afx limb stent graft.Approximately three (3) months later, a non-device related reintervention was performed with the implant of an afx limb stent graft and two (2) ovation ix iliac limbs.Now, approximately one (1) year later, a possible type ia endoleak was identified at routine follow-up.Reportedly, the physician planned to treat the patient with a non-endologix device on 06/13/2022.It is unknown if the reintervention took place.The patient status is not known.Clarification: the patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft, an afx vela infrarenal, an afx limb stent graft, and an ovation ix extender.This procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately (4) months post-initial procedure, a non-device related reintervention was performed with the implant of one (1) ovation ix iliac limb and two (2) afx limb stent grafts.Approximately seven (7) months later, a non-device related reintervention was performed with the implant of an ovation ix iliac limb and an afx limb stent graft.Approximately two (2) months later, a non-device related reintervention was performed with the implant of an afx limb stent graft and three (3) ovation ix iliac limbs.Now, approximately one (1) year later, a possible type ia endoleak was identified at routine follow-up.Reportedly, the physician planned to treat the patient with a non-endologix device on 06/13/2022.It is unknown if the reintervention took place.The patient status is not known.Additional information: clinical assessment confirmed that the patient was treated (as planned) on 13 june 2022.Reportedly, the patient was discharged home two days postoperative.Additional clinical assessment confirmed that the left iliac artery was poorly perfused (classified as stenosis).
|
| |
|