• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. INSERTER THREADED SHAFT; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. INSERTER THREADED SHAFT; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 2112-90009
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that a capri inserter, capri inserter inner shaft, and capri non-tapered driver were damaged during surgery.The surgeon malleted the inserter into the patient which caused deformation on the proximal end of the capri inserter (mallet marks) and deformation on the distal end of the capri inserter inner shaft.Images provided indicate the distal end of the capri non-tapered driver is also deformed.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.This record captures the capri inserter inner shaft.
 
Event or Problem Description
A company representative reported that a capri inserter, capri inserter inner shaft, and capri non-tapered driver were damaged during surgery.The surgeon malleted the inserter into the patient which caused deformation on the proximal end of the capri inserter (mallet marks) and deformation on the distal end of the capri inserter inner shaft.Images provided indicate the distal end of the capri non-tapered driver is also deformed.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.This record captures the capri inserter inner shaft.
 
Additional Manufacturer Narrative
A visual inspection was performed on the device and the reported issue of inserter damage and inserter inner shaft damage was confirmed.The proximal end of the inserter and the inserter inner shaft exhibits severe mallet marks indicating that excessive force was applied to these devices during implant insertion.A review of complaint history associated with the subject catalog and lot number was performed, and no adverse trend was observed.The device was inspected prior to the procedure and no damage was noticed.Hence, the cause of the reported event is application of excessive force by the user causing device deformation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTER THREADED SHAFT
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14780540
Report Number3004774118-2022-00250
Device Sequence Number5936413
Product Code MQP
UDI-Device Identifier10888857033665
UDI-Public10888857033665
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2112-90009
Device Catalogue Number2112-90009
Device Lot NumberKHCG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1