Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation ¿ evaluation on 13apr2022, cook india received a complaint from dr (b)(6), a representative at the (b)(6) hospital, located in the city of coimbatore tamil nadu india.The supplied stiffening cannula of the ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter, (rpn: ult7.0-35-25-p-5s-cldm-hc, lot: 10050663) broke into two pieces when removing.The drainage catheter and stiffener were removed as one and replaced with a new 7fr catheter set.The patient did not experience any adverse effects due to this occurrence.Reviews of the documentation, including the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, quality control procedures, and specifications, as well as a visual inspection of the returned device, were conducted during the investigation.The complaint device was returned for evaluation in a used and damaged condition.Upon removal of the metal stiffening cannula from the catheter, separation was confirmed.Both ends of the metal cannula's point of separation exhibited annealing.A bow was noted at the midpoint of the cannula.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are currently in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for lot 10050663 found no relevant nonconformances that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with this lot number.Evidence gathered upon review of the dmr, dhr, and device failure analysis, does not indicate the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Cook also reviewed product labeling.The product ifu, [t_multi2_rev1] ¿multipurpose drainage catheter,¿ provides the following information to the user related to the reported failure mode: how supplied: ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the information provided, examination of the returned product and the results of our investigation, it was concluded that a component failure unrelated to manufacturing or design deficiencies contributed to the event.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|