• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR HIP; EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 40H

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR HIP; EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 40H Back to Search Results
Model Number 942-01-40H
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this complaint was reported as wrong implant selection during surgery.The healthcare professional indicated that this event occurred during surgery, away from the patient.Significant adverse event was reported by the surgeon.There was a ten (10) minute delay in surgery but the surgery was completed as intended.The agent was present during surgery and did not have another suitable device available.The devices were returned to manufacturer and evaluated by registered medical assistant (rma) djo surgical.A review of the device history records indicate: twenty (20) pieces of 942-01-40h, acetabular liner, lot 629z1085,were processed on work order a13211163.These parts went through the sterilizer under sterilization run e220406aaa.Twenty (20) pieces of 341-14-709, knee insert, lot 132t1085, were processed on work order a13211380.These parts went through the sterilizer under sterilization run e220406aaa.Both work orders were processed by the same operator.The label on the bottom tray matches the work order and packaged part; however, the label on the front of the tray and outer packaging were swapped.A review of the complaint database indicates one other instance occurring for the same product and lot numbers.The surgery took place on the same day.The root cause of this complaint is likely due to the product being switched after the sterilization process.Hhe-2022-00007 was opened to capture the containment process for this issue.Rma examination: the reported devices were returned to djo surgical for evaluation.It is confirmed that the packaging for the acetabular liner contained a empower 3d knee insert.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Complaint - during surgery, two hip liner implant boxes were consecutively opened to find empower 3d knee inserts in the boxes.Surgeon wanted to use a 40mm, 10 degree liner, but was unable.Surgeon had to implant the 40mm non-hooded liner instead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR HIP
Common Device Name
EMPOWR ACETABULAR SYSTEM, LINER, 10° HOODED, HXE+, 40H
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14780967
Report Number1644408-2022-00813
Device Sequence Number14979471
Product Code LPH
UDI-Device Identifier00190446309833
UDI-Public(01)00190446309833
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number942-01-40H
Device Catalogue Number942-01-40H
Device Lot Number629Z1085
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Initial Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
341-14-708 LOT 132T1085; 341-14-708 LOT 132T1085
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexMale
-
-