• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. GXL NUETRAL LINER, G0 28MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. GXL NUETRAL LINER, G0 28MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number GXL NUETRAL LINER, G0 28MM ID
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by the legal brief, the female patient underwent a left total hip arthroplasty surgical procedure on (b)(6) 2013.Patient had a revision surgery performed on (b)(6) 2020 due to "catastrophic poly wear" and "metallosis.".
 
Additional Manufacturer Narrative
After further review of additional information received the following sections d1, d4, d10, g3, g4, g6, h2, h3, h4, h6, h7 and h9 have been updated accordingly.Section d10: concomitant product - 12/14 zirconia head 28mm +0mm neck (cat# 148-28-00 / serial# (b)(6)) (h3) based on the available information, the revision reported may have been the result of a combination of the risk factors including but not limited to use error, implant positioning, implant size selection, and patient factors, which led to polyethylene wear, metallosis, and osteolysis.However, this cannot be confirmed as the devices, radiographs, and operative notes were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GXL NUETRAL LINER, G0 28MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14781731
Report Number1038671-2022-00710
Device Sequence Number6108473
Product Code JDI
UDI-Device Identifier10885862022059
UDI-Public10885862022059
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K100269
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2017
Device Model NumberGXL NUETRAL LINER, G0 28MM ID
Device Catalogue Number130-28-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1731-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-