| Model Number |
3100 A |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the 3100a ventilator stopped running while on patient.At this time, there was no information regarding patient harm and intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: vyaire medical was able to confirm the issue - unit stopped running.Vyaire field service representative (fsr) evaluated the device on-site, and examined the unit for any faults.All voltages were with in specifications but pressures were needed to be recalibrated.Unit needs a 7 year pm (preventive maintenance).Unit will be operational following a pm.
|
| |
|