• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTIC BRONCHOSCOPE 3.2C 6.2TP 600L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTIC BRONCHOSCOPE 3.2C 6.2TP 600L Back to Search Results
Model Number FB-19TV
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Fiber image failure(broken).
 
Additional Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.We checked the returned unit and confirmed that the image guide fiber bundle (cfb) was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the image guide fiber bundle (cfb).In addition, we confirmed that the segment fluid damage, the image guide fiber bundle (cfb) fluid damage, the objective lens unit fluid damage, the lcb distal cover glass fluid damage, the operation channel buckled, the insertion flexible tube (ift) leak, the light guide cable leak, and the root brace rubber cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
FIBEROPTIC BRONCHOSCOPE 3.2C 6.2TP 600L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key14784357
Report Number9610877-2022-54909
Device Sequence Number14714926
Product Code EOQ
UDI-Device Identifier04961333083534
UDI-Public04961333083534
Combination Product (Y/N)N
Initial Reporter CountryLH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFB-19TV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Initial Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-