This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.We checked the returned unit and confirmed that the image guide fiber bundle (cfb) was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the image guide fiber bundle (cfb).In addition, we confirmed that the segment fluid damage, the image guide fiber bundle (cfb) fluid damage, the objective lens unit fluid damage, the lcb distal cover glass fluid damage, the operation channel buckled, the insertion flexible tube (ift) leak, the light guide cable leak, and the root brace rubber cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|