• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MONOJCT 20ML SYR L-LOCK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MONOJCT 20ML SYR L-LOCK; SYRINGE, PISTON Back to Search Results
Model Number 1182000777
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the inner piece is breaking off at the luer-lock end.The threads stay but the inner straight piece breaks off.They were losing costly medications at the same time and had to dispose of the whole syringe and medication.Per additional information received on 01jun2022, there was no medical intervention required.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.During the manufacturing process, a sampling of devices was visually inspected and physically tested with no issues recorded relating to the customer report.The product was manufactured between august 4th and 5th, 2022.The manufacturing site received one opened unit carton with 4 unsealed softpack packages with syringe assemblies included and 7 sealed softpack packages with syringe assemblies included.A total of 28 syringe assemblies were received for analysis.A complete investigation was performed.Visual inspection to the quality inspection standard was conducted on the samples and no issues were found relating to the reported condition.Control mechanisms are in place to prevent the occurrence and acceptance of the reported condition during molding, printing, assembly, and packaging processes, and to ensure components and finished product meet all quality inspection standards during the syringe assembly processes.Per procedure, complaint trends are evaluated during a monthly corrective and preventative action (capa) meeting to determine if a capa is warranted.At this time, there is not enough information and a capa will not be initiated.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOJCT 20ML SYR L-LOCK
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
1222 sherwood rd
norfolk NE 68701
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14786392
Report Number1915484-2022-01279
Device Sequence Number5058549
Product Code FMF
UDI-Device Identifier10884521010802
UDI-Public10884521010802
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1182000777
Device Catalogue Number1182000777
Device Lot Number122120X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date06/23/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-