U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL SOFT STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEM; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL SOFT STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEM; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00555860
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was contaminated and will not be returned for analysis; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a wallflex duodenal stent was to be implanted in the first part of the duodenum to treat a 10 mm over 2 cm malignant duodenal stricture during a stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous.During the procedure, when the wallflex duodenal stent delivery system was deployed halfway, there was no movement of the outer sheath that releases the stent.Then the wallflex duodenal stent suddenly deployed prematurely.The stent was removed from the patient with the use of rescue combo forceps.Another wallflex duodenal stent was implanted to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL SOFT STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEM
Common Device Name
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14786571
Report Number3005099803-2022-03310
Device Sequence Number5935565
Product Code MUM
UDI-Device Identifier08714729902492
UDI-Public08714729902492
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
SIMILAR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2023
Device Model NumberM00555860
Device Catalogue Number5586
Device Lot Number0027023859
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-