MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83786
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 26may2022.It was reported that the coil did not advance into the catheter.A 10mm x 20cm interlock-35 was selected for use.During the procedure, it was noted that the coil did not advance into the catheter.Also, the hairs on the coils were loose and deformed hairs.The coil and catheter were removed together from the patient's body and the procedure was completed with another of same device.No complications were reported and patient was stable post procedure.However, device analysis revealed that there were missing fiber bundles.
 
Additional Manufacturer Narrative
Initial reporter facility name: (b)(6).Device eval by mfr: the device was returned for analysis and evaluation was completed.Visual inspection revealed that the main coil was returned for this complaint.The main was kinked and stretched.No more damages were observed.Microscopic inspection revealed that the zap tip has a smooth surface.The interlocking arm was inspected, and no anomalies were noted.Dimensional inspection of the main coil revealed the components were within specification except the number of distal fiber bundles, which were only 19 out of 22 and the number of proximal fiber bundles, which were only 20 out of 22-32.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14787004
Report Number2134265-2022-06858
Device Sequence Number5876878
Product Code KRD
UDI-Device Identifier08714729793137
UDI-Public08714729793137
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K133208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/07/2024
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0027630854
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Initial Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
-
-