• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW INSTRUMENT; REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW INSTRUMENT; REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM Back to Search Results
Catalog Number NAT-024
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This investigation is still in progress.Once the investigation is complete a supplemental report will be provided.
 
Event or Problem Description
The customer reported the id now covid-19 instrument to be smoking after a hard reboot.There was no visible fire or any harm due to the smoke.The id now covid-19 instrument would no longer power on.There was no patient involvement.
 
Additional Manufacturer Narrative
Investigation report: the device was returned in a condition that made analysis impossible.The release documentation for instrument part number nat024 and instrument serial number (b)(6) was reviewed and the instrument met all release criteria.Notwithstanding, a review of complaints' did not reveal any unexpected results.In conclusion, the customer's returned instrument could not be verified for overheating when tested at abbott diagnostics scarborough.Therefore, a root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ID NOW INSTRUMENT
Common Device Name
REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
MDR Report Key14787162
Report Number1221359-2022-03025
Device Sequence Number4832182
Product Code OOI
UDI-Device Identifier00811877010593
UDI-Public0100811877010593
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K173653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNAT-024
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date06/23/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-