COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NGE-017115
Device Problems Break (1069); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) number = exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
It was reported that a user opened the packaging to a nforce nitinol helical stone extractor and found that the sheath near the handle was broken and the basket would not open.A new device was used to complete the procedure.There was no impact to the patient.
 
Event Description
The device was an ngage nitinol stone extractor and not a nforce nitinol helical stone extractor as originally reported in the previous medwatch report that was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Corrected information: b5, d1, d4 event description: it was reported that a user opened the packaging to a ngage nitinol stone extractor and found that the sheath near the handle was broken and the basket would not open.A new device was used to complete the procedure.There was no impact to the patient.Investigation ¿ evaluation a visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), manufacturing instructions, the instructions for use (ifu), and quality control procedures.One ngage nitinol stone extractor was returned for investigation in an open package with label.The basket sheath was kinked.The basket sheath and support sheath were no longer secured together.A functional test determined the handle did not actuate basket formation.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted.All extractors are verified to assure the basket opens and closes properly.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: suggested handling instructions for extractors and forceps important: excessive force could damage device.It is likely the basket sheath was initially kinked, which prevented the basket assembly from moving freely within the basket sheath, then the increased force between those two components caused the basket sheath and support sheath to separate.The cause for the damage is unknown.Excessive force may have been inadvertently applied to the device, however no information was known regarding device handling, therefore the cause of the issue could not be conclusively determined.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14787616
MDR Text Key302583519
Report Number1820334-2022-01110
Device Sequence Number1
Product Code FFL
UDI-Device Identifier10827002482941
UDI-Public(01)10827002482941(17)250112(10)14468465
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNGE-017115
Device Lot Number14468465
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received08/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-