MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT5
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An inventory management specialist reported that during an intraocular lens (iol) implant procedure, a torn lens was noted with patient contact.Additional information was received stating there was no patient harm and surgery completed on same day.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol returned in two pieces (cut in half) in the lens case.Solution and blood are dried on both surfaces of the optic and haptics.One haptic is intact, the other haptic is folded and adhered to the optic with solution.The optic is cut in two and scratched/marked-rejectable.The complainant indicates the use of viscoelastic, which is not qualified for use with associated cartridge.While we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14787705
Report Number9612169-2022-00293
Device Sequence Number20317168
Product Code MJP
UDI-Device Identifier00380652250856
UDI-Public00380652250856
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberSN6AT5
Device Catalogue NumberSN6AT5.130
Device Lot Number21204239
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/14/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date06/23/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE D; PROVISC; UNSPECIFIED INJECTOR
Patient SexUnknown
-
-