Model Number 1804400-18
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It is unknown if the device is returning for analysis.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a 75% stenosed, moderately calcified and moderately tortuous vessel in the right coronary artery (rca).A 3.5mm xience stent delivery system (sds) was deployed in the mid rca, however, the 4.0x18mm xience skypoint met resistance with the anatomy during advancement and failed to cross the lesion in the proximal rca.The distal edge of the stent was noted to be flared.Resistance was met with the lesion calcification during removal.There was no adverse patient effect.There was no report of clinically significant delay in the procedure.No additional information was provided.
 
Event or Problem Description
It was reported that the procedure was to treat a 75% stenosed, moderately calcified and moderately tortuous vessel in the right coronary artery (rca).A 3.5mm xience stent delivery system (sds) was deployed in the mid rca, however, the 4.0x18mm xience skypoint met resistance with the anatomy during advancement and failed to cross the lesion in the proximal rca.The distal edge of the stent was noted to be flared.Resistance was met with the lesion calcification during removal.There was no adverse patient effect.There was no report of clinically significant delay in the procedure.Subsequent to the initially filed report, the following information was provided: the sds was removed independently.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14788509
Report Number2024168-2022-06755
Device Sequence Number5058635
Product Code NIQ
UDI-Device Identifier08717648233487
UDI-Public08717648233487
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2024
Device Model Number1804400-18
Device Catalogue Number1804400-18
Device Lot Number108304A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer07/07/2022
Initial Report FDA Received Date06/23/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XIENCE STENT
Patient SexUnknown
-
-