Model Number 0998-00-0800-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had the arterial pressure and augmentation waveforms spike in size rate when on battery, but while on ac it returns back to normal.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) troubleshot, found issue was related to the front end board, ordered part and returned at a later date for repair.Installed a new front end board, and recalibrated pressure values.Safety, and functionality checks completed per the service manual and passed to factory specifications.Returned for clinical service upon completion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14788921
MDR Text Key301221092
Report Number2249723-2022-01470
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2022
Initial Date FDA Received06/23/2022
Supplement Dates Manufacturer Received03/05/2024
Supplement Dates FDA Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-