| Model Number |
165-001 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Unspecified Musculoskeletal problem (4535)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It is unknown why the failure was caused either surgeon technique, patient post-operative activise could have reported to the event.The failure of the tulip head was flared out and peeled where the bone screw's spherical head were forced through the taper lock.
|
| |
|
Event or Problem Description
|
|
It was reported that two centerline modular heads came off of the screws during surgery.
|
| |
|