| Model Number |
EXPRESS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker performed an initial evaluation of the customer's device and verified the reported issue.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device displayed all three icons (charge-pak, attention, and service wrench).The illumination of all three icons indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker further evaluated the customer's device.Stryker observed rust buildup or corrosion on several internal components.Stryker determined that the cause of the reported issue was due to contamination.The customer was provided with a replacement device.The returned device was archived by stryker.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device displayed all three icons (charge-pak, attention, and service wrench).The illumination of all three icons indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with this event.
|
| |
|