MEDELA LLC PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN); PUMP, BREAST, POWERED Back to Search Results
Model Number 101041361
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, the customer alleged to medela llc that her pump in style max flow had a milk backup.Additionally she alleged that she had mastitis.
 
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer, verifying using correct kit components; verifying parts are being washed according to instructions for use; verifying the customer did not have a heavy letdown and correct positioning being used while pumping.The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.In follow up with a complaint handler on (b)(6) 2022, the customer stated that she developed mastitis on (b)(6) 2022, and was prescribed antibiotics.Additionally she stated that the new pump is working great.Based on the results of our internal investigation (reference number (b)(4)),it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of (b)(6) 2013 to (b)(6) 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Additional Manufacturer Narrative
The device was returned with the customer's tubing and was evaluated on (b)(6) 2022.The device was tested with a lab kit (connectors and membranes) and the customer's tubing.Human milk residue was found on the aggregate (internal motor & pump assembly).After cleaning the aggregate, the vacuum reading were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN)
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
dave kurudza
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key14792062
Report Number1419937-2022-00057
Device Sequence Number5812542
Product Code HGX
UDI-Device Identifier00020451401454
UDI-Public020451401454
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101041361
Device Catalogue Number101041361
Device Lot Number0000039607
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Distributor Facility Aware Date06/16/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date06/23/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/03/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-