| Model Number |
115-150 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924); Seroma (2069)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported left side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record is not available due to insufficient device data.No contact information was provided for the initial reporter, therefore additional event, product, and/or patient details are not attainable.The reason for reoperation is: "rupture".
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The events of capsular contracture and seroma- late are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported the events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Patient reported left side rupture.The events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: one opening observed on posterior side assessed as fold flaw opening.Seroma-late: unable to observe as it is a medical event not related to the device.Capsular contracture: unable to observe as it is a medical event not related to the device.Double capsule: unable to observe as it is a medical event not related to the device.Crease/folding of implant: creases were observed.Malposition: unable to observe as it is a medical event not related to the device.Lump/nodule-ndr: not applicable as the event is not related to the device cyst-ndr: not applicable as the event is not related to the device.Additional observations: creases were observed.Wear abrasion observed.Red particles observed on the surface of the shell.Stress marks observed.
|
| |
|