| Model Number |
1804300-28 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the left anterior descending coronary artery that was heavily calcified and 90% stenosed.The 3.0x28mm xience skypoint stent system was advanced but met resistance and failed to reach the lesion due to the anatomy.During removal, resistance was noted.There was no reported adverse patient effect and no clinically significant delay in the procedure.A drug coated balloon was used to complete the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|