W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
PLC141400 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
08/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional devices implanted and/or related to this event: pll161407/(b)(4) udi (b)(4).Patient medications: aspirin, clopidogrel, heparin, acetaminophen and oxycodone.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore through the pivotal study for the gore® excluder® thoracoabdominal branch endoprosthesis (tambe device): on (b)(6) 2020, the patient was implanted with two gore® contralateral leg endoprostheses, a gore® iliac branch endoprosthesis and the gore® excluder® thoracoabdominal branch endoprosthesis (tambe device).Two gore® dryseal flex introducer sheath¿s were used for the procedure.The patient was being treated for an aortic aneurysm involving the visceral vessel(s).It was reported that on (b)(6) 2020, images revealed the patient had a distal left common iliac dissection.It is unknown which gore® dryseal flex introducer sheath was used in the left common iliac artery.Further investigation is being conducted.The adverse event is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: added component code g04122.H6: updated investigation findings code from 3233 to c19, a review of the manufacturing records indicated the lot met pre-release specifications.H6: updated investigation conclusion code from 11 to d15.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), potential device or procedure-related adverse events that may occur and/or require intervention include, but are not limited to, dissection, perforation, or ruptures of the aortic vessel and surrounding vasculature.
|
| |
|