• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL VITAL SIGNS¿ CABLE FOR CAM14 FOR MAC5500¿, 1.3M / 4.2FT; CABLE, ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL VITAL SIGNS¿ CABLE FOR CAM14 FOR MAC5500¿, 1.3M / 4.2FT; CABLE, ELECTRODE Back to Search Results
Model Number VITAL SIGNS¿ CABLE FOR CAM14 FOR MAC5500¿, 1.3M / 4.2FT
Medical Device Problem Code Fire (1245)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The customer reported that the vital signs¿ cable for cam14 for mac5500¿, 1.3m / 4.2ft locking pin is snapped.Due to this, it burned the power supply assembly mac 5000 - rohs and damaged the power supply cord ra control europe 10a 250v 2.5m.They also noticed the cracked writer base mac series and weak gas shock absorber cylinder.All of the damaged equipment have been replaced, tested for operation and leakage current.There is no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SIGNS¿ CABLE FOR CAM14 FOR MAC5500¿, 1.3M / 4.2FT
Common Device Name
CABLE, ELECTRODE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine CA 92618
Manufacturer (Section G)
CARLISLE MEDICAL TECHNOLOGIES (DONGGUAN) CO., LTD
no.2 xinhu industrial park,
dengwu village, qiaotou town
dongguan, 52353 3
CH   523533
Manufacturer Contact
erika bonilla
510 technology dr
irvine, CA 92618
MDR Report Key14795141
Report Number3013596024-2022-00005
Device Sequence Number14716922
Product Code IKD
UDI-Device Identifier10190752113541
UDI-Public(01)10190752113541
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVITAL SIGNS¿ CABLE FOR CAM14 FOR MAC5500¿, 1.3M / 4.2FT
Device Catalogue Number2016560-001
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2022
Initial Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-