W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA073902A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2019 the patient underwent endovascular treatment of a thoracoabdominal aortic aneurysm using a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) and gore® viabahn® vbx balloon expandable endoprostheses (vbx) in the (b)(6) tambe pivotal study.On (b)(6) 2019 a type ic endoleak was observed in the patient's right renal artery.An intervention was performed on (b)(6) 2019 whereby an additional vbx device was placed distally to extend the most distal vbx device in the patient's right renal artery.There were no further reported issues.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
C.1.Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2019 an intervention took place whereby balloon angioplasty was performed and an additional vbx device was used to extend the most distal vbx device in the patient's right renal artery.
 
Additional Manufacturer Narrative
B.5.Description of event - additional information added (balloon angioplasty performed) and original information clarified (wording).H.6.Investigation findings: code 213 - the review of the manufacturing paperwork verified that the lots involved in this event met all pre-release specifications.According to the gore® viabahn® vbx balloon expandable endoprosthesis instructions for use, as with all procedures that utilize techniques for introducing a catheter into a vessel, complications may be expected.These complications include, but are not limited to, device malapposition.H.6.Investigation findings: code 3233 updated to code 213 h.6.Investigation conclusions: code 11 updated to code 4315.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
jenna lopez
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14796262
Report Number2017233-2022-03050
Device Sequence Number5778921
Product Code PRL
UDI-Device Identifier00733132637270
UDI-Public00733132637270
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/29/2022
Device Model NumberBXA073902A
Device Catalogue NumberBXA073902A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN 81MG, CLOPIDOGREL 75MG
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
Patient Weight123 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-