• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 5; HIP STEM CEMENTLESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 5; HIP STEM CEMENTLESS Back to Search Results
Model Number 01.12.035
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
During surgery a quadra-h size 5 (ref.01.12.035, lot 152232b) was opened and in the package there was a quadra-s size 5 (ref.01.12.015, lot 101086a).A back-up quadra-h was used to complete the surgery.
 
Additional Manufacturer Narrative
Batch review performed on 23.Jun.2022.Lot 152232b: (b)(4).Expiration date: 2025-08-18.No anomalies found related to the problem.The other device involved in the mix-up: stem: quadra-s 01.12.015 cementless, sand blasted lat stem size 5 (k072857) lot 101086a: (b)(4).Expiration date: 2025-08-18.No anomalies found related to the problem.Analysis results: the mix-up has been confirmed; a box labeled with the ref.H 01.12.035 (lot 152232b, cementless, ha coated lat stem size 5) was found to include ref.01.12.015 (lot 101086a, quadra-s cementless, sand blasted lat stem size 5) and vice-versa.The letters "a" or "b" are present at the end of the lot numbers because they are assigned during the re-sterilization processing of the lots, as per medacta international internal procedures.Both lots involved in the action are composed of 1 device each.The corresponding item labeled as ref.01.12.015, lot 101086a, has been located and quarantined.It was under warehouse control and had not been distributed to a hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 5
Common Device Name
HIP STEM CEMENTLESS
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14797399
Report Number3005180920-2022-00488
Device Sequence Number4999810
Product Code JDI
UDI-Device Identifier07630030802171
UDI-Public07630030802171
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2025
Device Model Number01.12.035
Device Catalogue Number01.12.035
Device Lot Number152232B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2022
Initial Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient SexMale
Patient Weight80 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-