Model Number 60SP055
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use, the patient removed the tracheostomy tube.No patient injury was reported.
 
Additional Manufacturer Narrative
H10: no lot number was provided; therefore, device history record review could not be performed.A product sample and one photo was received for evaluation.Visual and functional testing was performed.Visual inspection found no defects were found in sample returned.Functional testing found when the sample was inflated with air, the cuff only inflated one side.According to the ifu the pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the time the cuff inflated completely and symmetrically, based on analysis the complaint is not confirmed.Device passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA UNCUFFED PEDIATRIC TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
4-1-5, izumi-chuo, izumi-ku,
minneapolis, MN 55442
MDR Report Key14798742
Report Number3012307300-2022-12436
Device Sequence Number12289235
Product Code JOH
UDI-Device Identifier15021312000436
UDI-Public15021312000436
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K912469
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60SP055
Device Catalogue Number60SP055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1