• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND TTS PEDIATRIC TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND TTS PEDIATRIC TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFS40
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Airway Obstruction (1699); Bradycardia (1751); Cyanosis (1798); Hypoxia (1918)
Date of Event 04/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the child had a history of ondine's syndrome and was ventilated during sleep via a tracheostomy tube.The patient was admitted to the department for a re-evaluation.At 8:30 in the morning, the professionals heard the ventilator alarm and observed that the ventilator was disconnected from the tracheostomy tube.The child was non reactive with blue lips, heart rate 37 and 7 percent oxygen saturation (spo2).Resuscitation was initiated.Ventilation was performed with a one-way valve self-filling balloon until help arrived.The cannula was blocked and suction was not possible.The cannula was changed and recovery of 100 percent spo2 and 80 cf occurred quickly with some insufflations.Additional information received indicated that the patient's status was recovered.The institution analyzed the situation and found that the main contributing factor was the lack of reporting of alarms from the patient's ventilator, which did not allow for rapid identification of the incident.
 
Additional Manufacturer Narrative
Other, other text: h4: device manufacturing date and d4: device expiration date could not be determined d4: device serial number/lot number is unknown h6: health effects - health impact: f27 h6: event problem and evaluation codes: update not required h10: no lot number was provided; therefore, device history record review could not be performed.No product sample nor photos were received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4)., corrected data: e2: health professional: no; e3: occupation: non-healthcare professional; e4 is unknown, no information provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA FLEXTEND TTS PEDIATRIC TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14798746
Report Number3012307300-2022-12437
Device Sequence Number17669343
Product Code JOH
UDI-Device Identifier10351688518620
UDI-Public10351688518620
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67PFS40
Device Catalogue Number67PFS40
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer03/08/2023
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age3 YR
Patient SexFemale
Patient Weight15 KG
-
-