Manufacturer's investigation conclusion: review of the submitted log files confirmed low flow alarms.According to the account, the low flow events were caused by cardiac tamponade.A direct correlation between (b)(6) and the reported cardiac tamponade and clots could not be conclusively determined through this investigation.The submitted controller event log file captured transient low flow hazard alarms on (b)(6) 2022.The system appeared to operate as intended at the set speed for the duration of the file.The patient remains ongoing on heartmate 3 left ventricular assist system (lvas), serial number (b)(6).The relevant sections of the device history records for (b)(6) were reviewed and showed no deviation from manufacturing or quality assurance specifications.The heartmate (hm) 3 left ventricular assist system (lvas) instructions for use (ifu) is currently available.Section 1 of the ifu, ¿introduction¿, lists pericardial fluid collection and peripheral thromboembolic event as adverse events that may be associated with the use of the heartmate 3 left ventricular assist system.Section 1 also provides an explanation of all pump parameters, including flow.This section explains that pump flow is a calculated value that is estimated based on pump power.Section 4 ¿system monitor¿ provides information about the pump flow display and the low flow hazard alarm condition.This section states that the low flow hazard alarm will be triggered when the estimated pump flow is less than 2.5 lpm and explains that changes in patient conditions can result in low flow.Section 7 ¿alarms and troubleshooting¿ describes all alarm conditions, including the low flow hazard, as well as the appropriate actions associated with them.Section 6 of the ifu, ¿patient care and management¿, lists thromboembolisms as a potential late post-implant complication.This section, under ¿anticoagulation¿, provides information regarding the recommended anticoagulation regimen, including inr values, as well as suggested anticoagulation modifications.Heartmate 3 lvas patient handbook is currently available.Section 5 "alarms and troubleshooting" outlines all system controller alarms as well as how to respond to each alarm condition.This document instructs the user that in the event of a low flow hazard alarm, call your hospital contact immediately for diagnosis and instructions.No further information was provided.The manufacturer is closing the file on this event.
|