• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,MCOL MG,3.7MM,16M; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,MCOL MG,3.7MM,16M; DENTAL IMPLANT Back to Search Results
Model Number TSVMB16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); Sensitivity of Teeth (2427)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Date of event unknown / not provided.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
The doctor advised of bone/jaw fracture at implant area #15.Patent advised that the bone/ fracture is a result of chewing/torque of dental implant.Patient agrees to surgical extraction of the implant.Symptoms as a result of the event: sensitivity and pain.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: investigation type codes were added: 3331, 4109, 4110, and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions codes were added: 67.One imp, tsv, mcol mg,3.7mm,16m (tsvmb16) was returned for investigation.Visual evaluation of the as returned product identified signs of use but no apparent signs of malfunction.Functional testing could not be performed for the reported event (bone fracture) as it is a medical condition.Based on the evaluation, device malfunction has not occurred.Additionally, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported event or device.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimmer biomet.Dhr review for the lot (1252052) revealed no deviations nor non-conformances which could have caused or contributed to the reported event.Complaint history review was performed for the reported lot number (1252052) for similar events (complaint category keywords: medical: bone fracture) and no other complaint was identified.Functional testing could not be performed for the reported event (bone fracture) as it is a medical condition.Therefore, the reported event could not be recreated.A definitive root cause could not be determined.Based on the investigation, risk review and ifu, the probable causes for the reported event are clinician error (improper surgical procedure) and patient factors (osteopenia).No further investigation or immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,MCOL MG,3.7MM,16M
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14799911
Report Number0002023141-2022-01615
Device Sequence Number5058433
Product Code DZE
UDI-Device Identifier00889024019751
UDI-Public(01)00889024019751(17)270209(10)1252052
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K111889
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberTSVMB16
Device Catalogue NumberTSVMB16
Device Lot Number1252052
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
Patient Weight77 KG
-
-