DAVOL INC., SUB. C.R. BARD, INC. -1213643 3DMAX MID; SURGICAL MESH Back to Search Results
Catalog Number 0116322
Medical Device Problem Code Expiration Date Error (2528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, an expired 3dmax mid was inadvertently implanted into the patient.The expiration date is located on multiple layers of the packaging.This event is confirmed as a use related error, with no malfunction of the device.Review of manufacturing records confirms product was manufactured to specification.To date, this is the only reported complaint from this manufacturing lot of (b)(4) units released for distribution in august 2020.
 
Event or Problem Description
As reported, during a robotic inguinal hernia repair procedure on (b)(6) 2022, a bard/davol 3dmax mid mesh which expired on 28-apr-2022 was implanted in a patient.It was reported that the mesh has been left implanted and the patient has not received any additional medical/surgical treatment.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DMAX MID
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14800908
Report Number1213643-2022-00476
Device Sequence Number4934997
Product Code FTL
UDI-Device Identifier00801741200793
UDI-Public(01)00801741200793
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K200818
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2022
Device Catalogue Number0116322
Device Lot NumberHUESAB15
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2022
Initial Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-