• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS INTERGARD WOVEN STRAIGHT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS INTERGARD WOVEN STRAIGHT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number IGW0028-30
Medical Device Problem Codes Use of Device Problem (1670); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to intervascular (complaint #(b)(4)) that during surgery, a graft (igw0028-30) was opened but its length appeared shorter than 30 cm.Therefore, another graft of the same product reference (igw0028-30) was added to the first one to complete the surgery.The surgery was significantly delayed.It was shared that there was a noticeable shortness when the products were opened during the surgery.There is no information about lot/serial number of the graft used and patient information.Since the allegation of shortness involves both implanted products, another mdr is being submitted regarding the second device that was implanted.Please note that due to this event, the stocks of the hospital were checked after the operation and another product of the same product reference, allegedly shorter than 30 cm, was identified.A complaint has been opened in order to handle the event (complaint #(b)(4)) and a mdr is being submitted for this case.
 
Additional Manufacturer Narrative
The device is not accessible as it remained implanted in the patient.The case was will be reviewed by the medical affairs.The medical review is pending.The investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Additional Manufacturer Narrative
Corrected data: in block b1, following investigation conclusions the field was changed from "adverse event & product problem" to "adverse event".In block h6, medical device problem code, code 2588 was changed to 1670.(4112/3221) the case was reviewed by the medical affairs whose assessment is: "this complaint describes an open surgical procedure requiring an intergard straight vascular graft.The complainant reported that the graft appeared to be shorter than the length stated on the packaging.It was implanted and then to complete the procedure and additional graft (of the same size and length) was implanted.The anastomosis to join the two grafts resulted in a prolonged procedure.The complainant has provided no additional information.It would be important to know the lot number of the devices that were implanted to further investigate if any similar devices were affected.Due to the lack of information provided with this complaint, it is not possible to determine the cause of the ¿shorter than specified length¿ of the grafts that were implanted." (4103/213) one product was returned to intervascular for examination (a product that seemed to present the same defect as the implanted products found in the stock inventory by the hospital ¿ see mfg#: 1640201-2022-00022).It was inspected by the quality assurance department, the conclusion is: "a length measurement of the product was performed as required in our applicable procedure.The length was found to be 31 cm for a specification of 30 cm (-0/+2 cm) which meets the expected length specification as mentioned in the labelling of the inspected product." (67) the conducted investigation indicates that the returned device was not defective.(19) no investigation on the manufacturing process of the products could be performed since no traceability data are available and the products remained implanted.However, based on the investigation findings of the related case (see mfg#: 1640201-2022-00022), we conclude that there was most probably also a visual misappreciation of the length of the products made by the observer and that the implanted products are meeting the dimension specifications as presented in the labelling.A letter will be sent to remind the user that the total graft length printed on the product labels is the minimum usable length.Indeed, according to our procedures and in accordance with the applicable international standard iso 7198, the usable length of vascular grafts is measured when graft is stretched at a medium tension without completely eliminating the embossing rings.
 
Event or Problem Description
See mfg initial reports#: 1640201-2022-00020 and #: 1640201-2022-00021.Complaint#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERGARD WOVEN STRAIGHT
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR  13705
Manufacturer (Section G)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR   13705
Manufacturer Contact
laurence richard
zone industrielle athelia i
la ciotat 13705
FR   13705
442084646
MDR Report Key14801055
Report Number1640201-2022-00020
Device Sequence Number4652882
Product Code DSY
UDI-Device Identifier00384401000242
UDI-Public(01)00384401000242
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K984294
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIGW0028-30
Device Catalogue NumberIGW0028-30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
-
-