Batch review performed on (b)(6) 2022.Lot 1910952b: 5 items manufactured and released on (b)(6) 2021.Expiration date: 2026-nov-22.No anomalies found related to the problem.To date, 2 items of the same lot have been sold with no similar reported event during the period of review.Original lot 1910952: 113 items manufactured and released on 03-jun-2020.Expiration date: 2025-may-24.No anomalies found related to the problem.To date, 103 items of the same lot have been sold with no similar reported event during the period of review.Clinical evaluation: partial revision 1 month after rsa due to joint dislocation.The reason of the adverse event is unknown.Liner was revised maintaining the same thickness of the liner (+6mm) but changing the ccd angle at 155° in order to have more stability.It may be possible that the dislocation occured due to laxity of the soft tissues.We do not see any reason to suspect a faulty device.Additional implant involved: batch review performed on (b)(6) 2022: reverse shoulder system 04.01.0121 humeral reverse hc liner ø36/+6mm (k170452) lot 2104818: 26 items manufactured and released on 23-jun-2021.Expiration date: 2026-jun-10.No anomalies found related to the problem.To date, 11 items of the same lot have been sold with no similar reported event during the period of review.
|