Model Number 07P9920
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely elevated alinity c creatinine results for one patient.The following data was provided: (b)(6) 2022 sid (b)(6) initial result = 2.88 mg/dl, (b)(6) 2022 sid (b)(6) initial result = 1.94 mg/dl, reference laboratory result = 1.9 mg/dl.No impact to patient management was reported.
 
Additional Manufacturer Narrative
Completed information for patient identifier: (b)(6).An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
This issue was previously reported under initial mdr number 3002809144-2022-00211.Upon further review, mdr number 3002809144-2022-00219 was a duplicate report.All further information will be documented under mdr number 3002809144-2022-00211.H3 other text : this issue was previously reported under initial mdr number 3002809144-2022-00211.Upon further review, mdr number 3002809144-2022-00219 was a duplicate report.All further information will be documented under mdr number 3002809144-2022-00211.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CREATININE REAGENT KIT
Common Device Name
ALKALINE PICRATE, COLORIMETRY, CREATININE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14801368
Report Number3002809144-2022-00219
Device Sequence Number17383696
Product Code CGX
UDI-Device Identifier00380740135102
UDI-Public00380740135102
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number07P9920
Device Catalogue Number07P99-20
Device Lot Number81084UN21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-