• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTANA MEDICAL SYSTEMS INC VENTANA HE 600 SYSTEM; SLIDE STAINER, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTANA MEDICAL SYSTEMS INC VENTANA HE 600 SYSTEM; SLIDE STAINER, AUTOMATED Back to Search Results
Catalog Number 06917259001
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated and is on-going.A follow-up report will be submitted upon the completion of the investigation.(b)(4).
 
Event or Problem Description
A customer from the united states reported that their he 600 system experienced fluid overflow in the middle staining module and fluid was seen coming out from where the cartridges load for the coverslip module.No alleged harm or injury.
 
Additional Manufacturer Narrative
Roche has confirmed complaints regarding fluid leaks within the middle stainer module of the ventana he 600 that likely resulted in electrical shorts at the plate heater connections that then escalated into fires.In the cases, the fire originated from the middle stainer modules at the backside of the connectors for the plate heaters.A third party fire analysis investigation was conducted for the cases and a capa has been initiated.Because the fire and any associated smoke could extend beyond the instrument, there is a risk of injury, potentially serious, from smoke inhalation and/or direct contact with the fire itself.Routine fire/smoke safety and evacuation procedures mitigate the risk of injury.Additionally, given the ability to retest the patient sample, the probability that this failure mode would result in adverse health consequences for patients is estimated as very low.No injuries have been reported to date.Consignees have been notified to not leave the instrument unattended when the power is on.Consignees were also instructed not to place the system in standby mode overnight when unattended until further mitigation actions have been implemented and to please turn off the power of the instrument when it is unattended.A stainer temperature monitoring update will be implemented and plate heater connectors will be coated with liquid electrical tape (let).These actions mitigate the risk by monitoring the temperature and by protecting the sources from leaks and subsequently a short circuit, which would prevent the possibility of heat generation.Both stainer temperature monitoring and liquid electrical tape updates are already implemented in the field in the us.An additional software update (v1.10.2 software release) to turn off the power to the stainer was made available on 14-oct-2022 and is scheduled to be installed on all instruments in the field by 14-oct-2023.This feature powers off the stainer whenever an excessive heater temperature is detected by the system.Component code as well as the eval result and conclusion codes were added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTANA HE 600 SYSTEM
Common Device Name
SLIDE STAINER, AUTOMATED
Manufacturer (Section D)
VENTANA MEDICAL SYSTEMS INC
1910 e innovation park dr
tuscon AZ 85755
Manufacturer Contact
stacie-ann creighton
1910 east innovation park dr
na
tuscon, AZ 85755
9082537112
MDR Report Key14801769
Report Number2028492-2022-00004
Device Sequence Number17354575
Product Code KPA
UDI-Device Identifier04015630976010
UDI-Public04015630976010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
510K EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification,Recall
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number06917259001
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date06/24/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2028492-06-01-2022-002-C
Patient Sequence Number1
Patient SexUnknown
-
-